Issue I509 [defect,prg,resolved-in-8.0-only]
  "illegal start for seg" error.

  The following error:

    SIMION Error: Runtime Error: Illegal Start for Time_Step Seg

  may occur when using a workbench with overlapping electric and
  magnetic potential arrays that each have a PRG program attached.

  A workaround may be to define a tstep_adjust segment in both
  the magnetic and electric programs.

  Reported by steveb-20080806.  Already fixed in 8.0.5-TEST20.

  Also reported in http://www.simion.com/discuss/viewtopic?p=2923
  (7.0).

  Affects 8.0.0-8.0.5-TEST20.
  Resolved in 8.0.5-TEST21.